• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 12, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios GOBlock System is a disposable, self contained infusion system.
Recall Number Z-1461-2013
Classification Class I
Code Info Pump Part 510613 , Lot 11-100815 - Lot 12-10101060 found in: GoBlock Kit 510612, Lot 11-100817 to 13-101475
Product Distributed Qty 33,543 all kits
Reason For Recall Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the 300 mL (dual reservoir), 2 mL/hr GOPump (PN510076). To date, there have been 5 complaints received, 2 of which involved serious injury. There have been no patient deaths reported. The risk of patient toxicity due to the rapid influx of medication include seizure, dysrhythmia, and death.

Event Detail

Event Id 65192
Product Type Devices
Status Ongoing
Recalling Firm Symbios Medical Products, LLC
City Indianapolis
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern nationwide USA,
-
-