Product Detail

Product Description Brainlab, Spine & Trauma 3D 2.0, Navigation Software. An intraoperative image-guided localization system to enable minimally invasive surgery. Catalog Number: 22264
Recall Number Z-1469-2014
Classification Class II
Code Info Catalog Number: 22264; Serial Numbers: 000000000038012001 , 000000000060912001 , 000000000189912001 , 000000000224912001 , 000000000235512001 , 000000000282912001 , 000000000313612001 , 000000000319412001 , 000000000351112001 , 000000000356912001 , 000000000367212001 , 000000000433712001 , 000000000482012001 , 000000000502812001 , 000000000645612001 , 000000000731112001 , 000000000837312001 , 000000000898312001 , 000000000956812001 , 000000000998412001 , 000000001012812001 , 000000001105412001 , 000000001231812001 , 000000001353512001 , 000000001355412001 , 000000001433712001 , 000000001473712001 , 000000001493512001 , 000000001493712001 , 000000001566212001 , 000000001579112001 , 000000001635112001 , 000000001758212001 , 000000001770212001 , 000000001789312001 , 000000001868312001 , 000000001895412001 , 000000001909912001 , 000000002036112001 , 000000002309812001 , 000000002439112001 , 000000002468312001 , 000000002503012001 , 000000002573012001 , 000000002682913001 , 000000003000913001 , 000000003053313001 , 000000003066813001 , 000000003071613001 , 000000003087613001 , 000000003109513001 , 000000003109613001 , 000000003230113001 , 000000003246713001 , 000000003304113001 , 000000003331913001 , 000000003379313001 , 000000003453413001 , 000000003453513001 , 000000003622213001 , 000000003671813001 , 000000003744413001 , 000000003758413001 , 000000003819613001 , 000000003824913001 , 000000003866013001 , 000000003937813001 , 000000004094013001 , 000000004147413001 , 000000004356413001 , 000000004404013001 , 000000004489313001 , 000000004489413001 , 000000004505313001 , 000000004821913001 , 000000004882013001 , 000000004905013001 , 000000005396313001 , 000000005407413001 , 000000005505113001 , 000000005506013001 , 000000005531913001 , 000000005552313001 , 000000005578713001 , 000000005659713001 , 000000005702513001 , 000000005788213001 , 000000005817113001 , 000000005904513001 , 000000005904613001 , 000000005904713001 , 000000005904813001 , 000000005946713001 , 000000006046413001 , 000000006046513001 , 000000006096913001 , 000000006136313001 , 000000006233113001 , 000000006296713001 , 000000006426913001 , 000000006427213001 , 000000006621613001 , 000000006621613002 , 000000006656613001 , 000000006792613001 , 000000006833413001 , 000000006839413001 , 000000006998813001 , 000000007000113001 , 000000007122113001 , 000000007172913001 , 000000007405013001 , 000000007588513001 , 000000007637114001 , 000000007679714001 , 000000007698714001 , 000000007784614001 , 000000007859414001 , 000000007906314001 , 000000008020214001 , 000000008031314001 , 000000008184514001 , 000000008193214001 , 000000008199814001 , 000000008214014001 , 000000008259514001 , 000000008383314001 , 000000009718312001 , 000000009720812001 , 000000009781112001 , 000000009802512001 , 000000009901612001 , 000000009983912001
Product Distributed Qty 133 units
Reason For Recall The "fluoro match registration" function of the Navigation Software Spine & Trauma 3D 2.0 allows the user to intraoperatively match e.g., CT data sets to the current patient anatomy visible on fluoroscopic images, used by the navigation software during spinal surgeries to display the position and orientation of instruments. For this registration function in combination with a digitally integrated C-arm, the software algorithm may not find an adequate match between the intraoperatively acquired 2D fluoroscopic images and the preoperative CT or other compatible 3D data sets. If such an incorrect match would occur, the display of instruments by the navigation system would be shifted compared to the actual patient anatomy.

Event Detail

Event Id 67569
Product Type Devices
Status Ongoing
Recalling Firm Brainlab AG
City Feldkirchen
State
Country DE
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-11-18
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US Distribution including the states of AZ, CA,GA, IL, IN, LA, MA, MI, NY, NC, OH, PA, TN, TX and WI and the countries of Australia, Austria, Belgium, Brazil, Chile, Czech Republic, France, Germany, Italy, Japan, Malaysia, Netherlands, Norway, Philippines, Poland, Russian Federation, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
-
-