Product Detail

Product Description DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.
Recall Number Z-1520-2013
Classification Class II
Code Info Part Number 230795000/Lot Number 5120443
Product Distributed Qty 20 units
Reason For Recall The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.

Event Detail

Event Id 65081
Product Type Devices
Status Ongoing
Recalling Firm DePuy Orthopaedics, Inc.
City Warsaw
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-03
Initial Firm Notification of Consignee or Public E-Mail
Distribution Pattern Worldwide Distribution - USA Nationwide and countries of: Australia, Canada, Czech Republic, England, Germany, Switzerland and Turkey.
-
-