• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 14, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad Laboratories, Inc. in vitro diagnostic. The Variant II Hemoglobin testing system uses the principals of high performance liquid chromatography (HPLC) for the separation and determination of normal and abnormal hemoglobin.
Recall Number Z-1537-2014
Classification Class II
Code Info Catalog number: 270-2103 (250 tests; Reorder pack lot number: 64005302, Expiration: 4/2015; Catalog number: 270-2154 (500 tests) Reorder pack lot number: 64005207, Expiration: 6/2015; 64004750, Expiration: 11/2015
Product Distributed Qty 457 kits
Reason For Recall There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.

Event Detail

Event Id 68016
Product Type Devices
Status Ongoing
Recalling Firm Bio-Rad Laboratories, Inc.
City Hercules
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-04-08
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distrubution ; USA INCLUDING CA, NY, GA, NC, AL, NJ, CT, WA, NY, LA, Il, MA and Internationally to China, Australia, Canada, France, Great Britain, Hong Kong, Thailand, Singapore,and Brazil.
-
-