• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 14, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Recall Number Z-1542-2014
Classification Class II
Code Info Serial Number 2000
Product Distributed Qty 1
Reason For Recall All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on the bed preventing bed high/low and contour functionality from occurring.

Event Detail

Event Id 67871
Product Type Devices
Status Ongoing
Recalling Firm Med-Mizer, Inc.
City Batesville
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-04-18
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern USA (nationwide)
-
-