Enforcement Report - Week of May 14, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Refill Kit 8551/856X (includes Models 8561, 8562 and 8564) for use with Medtronic Implantable Programmable Infusion Pumps. Product Usage: The Refill Kits are intended for use in refilling Medtronic programmable, implantable infusion pumps with the exception of Medtronic MiniMed infusion pumps.
Recall Number Z-1568-2014
Classification Class II
Code Info Lot numbers CS3643, CS3613, CS3663, N441998, N442081, N442131, N440259, N440527, N440582, N440654, N440909, N440978, N441199, N441268, N441806, N441817, N442178, N442513, N442532, N442560, N442588, N442831, and N443427. Affected lots were distributed between January 2014 and March 2014.
Product Distributed Qty 13,784
Reason For Recall Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem with the Lioresal Intrathecal (baclofen injection) contained in the affected drug refill kits. There have been no patient injuries beyond additional needle sticks associated with this issue.

Event Detail

Event Id 68031
Product Type Devices
Status Ongoing
Recalling Firm Medtronic Neuromodulation
City Minneapolis
State MN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-04-04
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Worldwide Distribution - USA Nationwide in the states of NM, ND, RI, and VT and the countries of Australia, Singapore, Thailand, Central and Eastern Europe, Western Europe, Middle East, & Panama.
-
-