Product Detail

Product Description MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.
Recall Number Z-1622-2013
Classification Class II
Code Info YUNO OTN Mobile Operating Table: 1433.01F0 Extension Device: 1433.62A1
Product Distributed Qty 13 Tables; 24 Extension devices
Reason For Recall MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released, skin which might be within the gap could be pinched or cut.

Event Detail

Event Id 65159
Product Type Devices
Status Ongoing
Recalling Firm Maquet Cardiovascular Us Sales, Llc
City Wayne
State NJ
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-04-26
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.
-
-