• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of May 28, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description syngo Imaging XS Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced postprocessing applications.
Recall Number Z-1622-2014
Classification Class II
Code Info material number: 10014063, with serial numbers: 1334 1065 1351 1491 1417 1269 1076 1043 1429 1353 1392 1258 1398 1445 1496 1279 1116 1090 1229 1080 1005 1081 1310 1246 1226 1057 1379 1396 1084 1322 1107 1453 1124 1343 1198 1432 1058 1247 1278 1082 1055 1049 1329 1203 1490 1083 1316 1395 1053 1079 1387 1448 1218 1451 1389 1386 1064 1500 1436 1074 1443 1210 1408 1427
Product Distributed Qty 64
Reason For Recall A complaint reported that the archiving method at one customer site was changed from implicit to explicit, and the cleanup procedure enabled by the explicit archivingmethod may have deleted data from archive itself.

Event Detail

Event Id 68107
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Malvern
State PA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-04-03
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution
-
-