Product Detail

Product Description CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535)
Recall Number Z-1634-2014
Classification Class II
Code Info Catalog Number or Part Number 145891 500AVHCT Serial Number: BPM SN 8015552 (installed into CDI 500 Monitor SN 3535)
Product Distributed Qty 1
Reason For Recall There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing acceptance criteria for a required drift test.

Event Detail

Event Id 68240
Product Type Devices
Status Ongoing
Recalling Firm Terumo Cardiovascular Systems Corporation
City Ann Arbor
State MI
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-05-09
Initial Firm Notification of Consignee or Public Telephone
Distribution Pattern one customer in IA
-
-