Product Detail

Product Description Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T
Recall Number Z-1693-2013
Classification Class II
Code Info Part Number: 640919
Product Distributed Qty 725 units
Reason For Recall Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

Event Detail

Event Id 65090
Product Type Devices
Status Ongoing
Recalling Firm Fresenius Medical Care Holdings, Inc.
City Waltham
State MA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-24
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.
-
-