Product Detail

Product Description Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.
Recall Number Z-1700-2014
Classification Class II
Code Info Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W
Product Distributed Qty 11 units
Reason For Recall Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Event Detail

Event Id 68298
Product Type Devices
Status Ongoing
Recalling Firm Hologic, Inc.
City Danbury
State CT
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-05-13
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
-
-