• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of July 17, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description REF 154210 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size 6ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.
Recall Number Z-1703-2013
Classification Class II
Code Info M375490, M444720 , M886400, M956080
Product Distributed Qty 22
Reason For Recall Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

Event Detail

Event Id 65459
Product Type Devices
Status Ongoing
Recalling Firm Biomet, Inc.
City Warsaw
State IN
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-06-07
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
-
-