• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of June 11, 2014

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.
Recall Number Z-1719-2014
Classification Class II
Code Info Beginning Lot Code 1204275; Ending Lot Code 1401075 (Lot code indicates the date of manufacture = yymmdds, where s is the shift).
Product Distributed Qty Domestic - 12,600 units, Foreign - 44,280 units
Reason For Recall ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.

Event Detail

Event Id 68193
Product Type Devices
Status Ongoing
Recalling Firm ConMed Corporation
City Utica
State NY
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-05-08
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and the countries of Australia, Belgium, Brazil, Canada, Ecuador, El Salvador, Finland, France, Great Britain, Honduras, India, Indonesia, Israel, Italy, Japan, Latvia, Lebanon, Malaysia, Mexico, Panama, Philippines, Portugal, Saudi Arabia, South Africa, Spain, Thailand, Trinidad, United Arab Emirates and Uruguay.
-
-