Product Detail

Product Description Siemens SOMATOM Emotion 16 and SOMATOM Spirit Computed Tomography Systems Product Usage: To produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
Recall Number Z-1721-2014
Classification Class II
Code Info Material numbers: 10165977, 10045692 with serial numbers: 32187 31158 31215 31222 31313 31354 32118 32120 32123 32125 32126 32127 32129 32139 32141 32156 32159
Product Distributed Qty 17
Reason For Recall The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.

Event Detail

Event Id 67967
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Malvern
State PA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2014-03-21
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern US Nationwide Distribution in the states of IA, TX, FL, NE, NY, OH, NC, KS, IL, CA, OK including PR.
-
-