Product Description Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Recall Number Z-1725-2014
Classification Class II
Code Info Leica Ceiling Mount MSF-1; Article Numbers: 10446781, 10446782
Product Distributed Qty 103 total
Reason For Recall The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Event Detail

Event Id 66274
Product Type Devices
Status Ongoing
Recalling Firm Leica Microsystems, Inc.
City Buffalo Grove
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-09-11
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
-
-