Product Description Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone lengthening, including conditions such as congenital mandibular deficiencies or post-traumatic defects. The Multi Vector Distractor can be utilized to perform bone transport procedures as an alternative to bone grafts and free flaps.
Recall Number Z-1745-2013
Classification Class II
Code Info Titanium Multi Vector Distractor Pin Holding Clamp, body, limited bone stock Part # 487.941 with lot numbers: 3769720, 3779569, 3788731, 3794434, 7806093, 7823463 and Titanium Multi Vector Distractor Pin Holding Clamp, ramus, limited bone stock Part # 487.942 with lot numbers: 3769706, 3779571, 3788741, 7806096, 7823464.
Product Distributed Qty 121
Reason For Recall Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however there is a remote potential that additional intervention would be required due to a non-functioning clamp.

Event Detail

Event Id 65561
Product Type Devices
Status Ongoing
Recalling Firm Synthes USA HQ, Inc.
City West Chester
State PA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-31
Initial Firm Notification of Consignee or Public Visit