• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of July 31, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.
Recall Number Z-1795-2013
Classification Class II
Code Info Model Number 5895003 with serial numbers 1052 through 2500.
Product Distributed Qty 401
Reason For Recall Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squeezed.

Event Detail

Event Id 65604
Product Type Devices
Status Ongoing
Recalling Firm Siemens Medical Solutions USA, Inc
City Malvern
State PA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-06-05
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide Distribution.
-
-