Product Detail

Product Description Leica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor for the following laboratory applications: fixation, dehydration, infiltration with intermedium, and the paraffin infiltration of histological tissue specimens.
Recall Number Z-1815-2013
Classification Class II
Code Info 1) Model ASP6025 120V/50-60 Hz; Serial Numbers: 171, 200, 202, 204, 206, 208, 210, 212, 214, 216, 218, 220, 222, 224, 226, 228, 230, 232, 234, 236, 238, 240, 242, 244, 246, 248, 250, 252, 254, 256, 258, 260, 262, 264, 266, 268, 270, 272, 274, 276, 278, 280, 282, 284, 286, 288, 290, 292, 294; 2) Model ASP6025 230V/50-60 Hz; Serial Numbers: 162, 163, 168, 169, 170, 183, 184, 201, 203, 205, 207, 209, 211, 213, 215, 217, 219, 221, 223, 225, 227, 229, 231, 233, 235, 237, 239, 241, 243, 245, 247, 249, 251, 253, 255, 257, 259, 261, 263, 265, 267, 269, 271, 273, 275, 277, 279, 281, 283, 285, 287, 289, 291, 293, 295, 297, 299, 301, 303, 305, 307, 309, 311, 313, 315, 317, 319, 320, 321, 323, 325, 327, 328, 329, 331, 333, 335, 337, 339
Product Distributed Qty 128 machines
Reason For Recall The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure failure of the device in combination with a bottle with incorrect dimensions, overflow can cause contamination of reagents. Additionally, the current instructions for use does not highlight the importance to check the filling levels of all reagent bottles (RTU and system bottles) prior to every process run. Without this important information, risk of tissue damage or loss is possible. Therefore the missing information will be distributed to all customers by a Field Service Notice.

Event Detail

Event Id 63935
Product Type Devices
Status Ongoing
Recalling Firm Leica Microsystems, Inc.
City Buffalo Grove
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-11-13
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and internationally to Australia, Belgium, Brazil, Chile, France, Germany, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, Philipines, Singapore, South Africa, Spain, Switzerland, Thailand, Turkey, and United Kingdom.
-
-