Product Detail

Product Description TX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator that emulsifies and removes soft tissue. The TX1 Console provides control over each mode of operation, as well as the user interface via LCD touchscreen and audible tones for confirmation of user selections. Indiciated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.
Recall Number Z-1817-2012
Classification Class II
Code Info Product Codes: LFL, GEI, Serial #: 55400001, 55400002, 55400003. Part Order No: 554-0034-000 510(K) Mp.: K101561
Product Distributed Qty 5units
Reason For Recall Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.

Event Detail

Event Id 61253
Product Type Devices
Status Terminated
Recalling Firm American Optisurgical Inc
City Lake Forest
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-02-22
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Nationwide distribution: MN only.
-
-