| Product Description | Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin, troponin I and B-type natriuretic peptide in EDTA anticoagulated whole blood and plasma specimens. |
|---|---|
| Recall Number | Z-1928-2012 |
| Classification | Class I |
| Code Info | Phase 1: W49604B, W49606B, W49607B, W49611B, W49612B, W49613B, W49614B, W49615B, W49616B, W49617B, W49620B, W49622B, W49623B, W49624B, W49625B, W49627B, W49628B, W49629B, W49630B, W49631B, W49633B, W49634B, W49635B, W49638B, W49639B, W50726B, W50727B, W50728B, W50731B, W50733B, W50734B, W50738B, and W50739B. Phase 2: W49609B, W49610B, W49618B, W49619B, W49626B, W49632B, W49640B, W50737B, and W50744B. |
| Product Distributed Qty | 16,962 kits |
| Reason For Recall | Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing. |
Event Detail
| Event Id | 61927 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Alere San Diego, Inc. |
| City | San Diego |
| State | CA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-05-22 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution |
-