Product Detail
| Product Description | Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. |
|---|---|
| Recall Number | Z-1961-2012 |
| Classification | Class II |
| Code Info | All serial/lot numbers |
| Product Distributed Qty | 6388 units |
| Reason For Recall | The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions. |
Event Detail
| Event Id | 62346 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Stryker Endoscopy |
| City | San Jose |
| State | CA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-19 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution--USA (nationwide) |
-