Product Detail

Product Description Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.
Recall Number Z-1961-2012
Classification Class II
Code Info All serial/lot numbers
Product Distributed Qty 6388 units
Reason For Recall The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.

Event Detail

Event Id 62346
Product Type Devices
Status Ongoing
Recalling Firm Stryker Endoscopy
City San Jose
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-19
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution--USA (nationwide)
-
-