• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 28, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Philips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage: MobileDiagnost wDR is a mobile digital radiography X-Ray System. It is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing, or lying in the prone or supine positions. Not intended for mammography.
Recall Number Z-1964-2013
Classification Class II
Code Info System Code - 712001
Product Distributed Qty 240 systems US -- 1 system within Puerto Rico
Reason For Recall The Instructions for Use (IFU) for the MobileDiagnost wDR fails to comply with a Federal standard. The IFU does not provide tolerance levels for several generator values as required by 1020.30(h)(3).

Event Detail

Event Id 65839
Product Type Devices
Status Ongoing
Recalling Firm Philips Healthcare Inc.
City Andover
State MA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-06-11
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern US Nationwide Distribution including Puerto Rico
-
-