Enforcement Report - Week of August 28, 2013

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description CHAMPION Medical Recliner, AC Powered with heat and massage, one chair per carton, or one chair wrapped in a blanket if delivered in person. Recliner provides heat and massage after surgery.
Recall Number Z-1980-2013
Classification Class II
Code Info The following model numbers are affected by this recall: 54PGS636-7JM; 56PLV04-TRL4LJM; 59PT09-F7CJM; 68PT96-TS7JM; 86PT26-TS7JM3; 54PLV04-TRL4LJM; 56PSR20-TS7JM; 59PT19-TFS7PJM3; 85PSR12-TS7JM3; 86PT45-T7KJM; 54PSR20-TS7JM; 56PT21-TS7JM; 65PEV943X-TS7JM3; 85PSR30-TS7JM3; 89PCHA06-TFS7HJM; 54PT26-TS7CJM3; 56PT26-TS7CJM3; 65PFS26-TS7JM; 85PSU103-TS7JM; 89PEV921X-TFS7HPJM3; 54PT45-NJM 56PT45-NJM 65PSH12-S7JM 85PT09-TS7JM3 89PFS04-TFS7HPJM3; 54PT45-TNJM; 56PT45-RNJM; 65PT09-TS7JM; 85PT19-T7JM; 89PFS13-TFS7HPJM3; 54PT45-TS7JM; 58PCHA06-TFS7HJM; 65PT21-TS7JM; 85PT26-TS7JM3; 89PFS15-TFS7HPJM; 54PT56-T7JM; 59PGS493-TFS7HPJM; 66PCHA04-TS7JM; 86PCHA06-TS7JM; 89PFS26-TFS7HPJM3; 54PT96-TS7JM3; 59PGS493-TFS7HPJM3; 66PEV923X-TS7JM3; 86PSU103-TS7JM; 89PFS32-TFS7HPJM3; 56PGS636-7JM; 59PSR12-FS7JM3; 66PFS26-TS7JM; and 86PT09-TS7JM3.
Product Distributed Qty 285 units
Reason For Recall There is a possibility of elevated temperatures in the seat pad. In addition, the hospital cord set does not agree with the labeling (information in manual).

Event Detail

Event Id 65552
Product Type Devices
Status Ongoing
Recalling Firm Invacare Corporation
City Elyria
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-06-18
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Nationwide distribution: USA including states of: CA, FL, IN, MN, NJ, NY, OH, RI, SC, SD, TX, and VA.
-
-