Product Detail

Product Description Intellispace Portal software. Intellispace Portal is a multimodality thin client applications server that delivers full diagnostic viewing and clinical applications to the enterprise.
Recall Number Z-1988-2012
Classification Class II
Code Info Model number 881001; Serial numbers: 81172, 79937, 85083, 81097, 85145, 79916, 81132, 85011, 82020, 85007, 85006, 85081, 81028, 80914, 82004, 85041, 81180, 82016, 30019, 80910, 81133, 81098, 81169, 79940, 80980, 82036, 82007, 85084, 85085, 82039, 81127, 81165, 30003, 80818, 30007, 85042, 81184, 82021, 82030, 82023, 82051, 82026, 82024, 82028, 82050, 82025, 82029, 82027, 30018, 79933, 82015, 85089, 30036, 81183, 82037, 80425, 81157, 85025, 82011, 81174, 80965, 80966, 81026, 35006, 82009, 81013, 82003 & 80891.
Product Distributed Qty 96
Reason For Recall Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that different patients have been selected.

Event Detail

Event Id 62107
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-23
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, FL, GA, IA, IN, MI, NC, NJ, NY, OK, PA, SC, TN, TX, and WA and the countries of Australia, Austria, Canada, China, Czech Republic, Denmark, and France.
-
-