Enforcement Report - Week of July 18, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. Intellispace Portal is a medical software system that allows multiple users to remotely access the system from compatible computers on a network.
Recall Number Z-2000-2012
Classification Class II
Code Info Model number: 881001; serial numbers: 80818, 85088, 81178, 82006, 82038, 81063, 82018, 79935, 81150, 81013, 81173, 85068, 79932, 79943, 80911, 82019, 81027, 85049, 85043, 82008, 81119, 81089, 85108, 85026, 85146, 80994, 81021, 81182, 81017, 85067, 81161, 81164, 79936, 85103, 81147, 85110, 85082, 81179, 81035, 85046, 81031, 81129, 81004 & 85104.
Product Distributed Qty 44
Reason For Recall In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.

Event Detail

Event Id 62128
Product Type Devices
Status Ongoing
Recalling Firm Philips Medical Systems (Cleveland) Inc
City Cleveland
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-22
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US (nationwide) including the states of: AR, LA, MI, NC, NY, OH, OK and PA and the countries of: Australia, Canada, Denmark, France, Germany, Israel, Netherlands, Spain and Switzerland.
-
-