Product Detail

Product Description Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Recall Number Z-2037-2012
Classification Class II
Code Info Serial #'s S5001 - S5011
Product Distributed Qty Six (6) units
Reason For Recall Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.

Event Detail

Event Id 62372
Product Type Devices
Status Ongoing
Recalling Firm Hitachi Medical Systems America Inc
City Twinsburg
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-05-04
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.
-
-