Enforcement Report - Week of August 1, 2012
-
Back to Previous Page
Product Detail
| Product Description | Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128 |
|---|---|
| Recall Number | Z-2079-2012 |
| Classification | Class II |
| Code Info | Serial numbers: W50024258; W50024259; W50026573; W50025790; W50017381; W50020296; W50028138; W50020952 |
| Product Distributed Qty | 8 washers |
| Reason For Recall | Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error. |
Event Detail
| Event Id | 61974 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Getinge USA Inc |
| City | Rochester |
| State | NY | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-05-25 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA including Maryland, Indiana, New Mexico, Massachuesetts, Texas, Kansas and the country of Canada |
-