Product Detail

Product Description UroView 2800; MDL number A349855. Product Usage: Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical application may include but are not limited to urologic and endoscopic procedures. The system may be ued for other imaging applications at the physician's discretion.
Recall Number Z-2127-2012
Classification Class II
Code Info 62-0212 , 69-0054 , 82-0023-RC , 82-0034 , 82-0072 , 82-0274 , 82-0587, 8S-0668 , 82-1710 , 82-1803 , 82-2147 , 82-2620 , 82-2754 , 82-2815 , 82-3014 , 82-3234 , 82-7010-MH , 86-0254 , 86-0354 , 86-1078 , 89-0505 , 89-0509 , 89-0595 , 89-0605 , 89-0987 , 89-1048 , 89-1776 , 89-1837 , 89-2132 , 89-2303 , 89-2335 , 89-2348 , 89-2350 , 89-2610 , 89-2645 , 8S-0066 , 8S-0079 , 8S-0341 , 8S-0562 , 8S-0615-C , 8S-0847 , 8S-0849-C , 8S-1312 , 8S-1971 , 8S-2015-N , 8S-2116 , 8S-2117 , 8S-2199 , 8S-2204 , 8S-2320 , 8S-2411 , P4-0097-R , P4-0101-R , P4-0325-R , P4-0346-L , P4-0477-L , P6-0432-R , W2-0012 , W2-0580-R , W2-0591-R , 82-0030
Product Distributed Qty Part of a total of 82 units.
Reason For Recall Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and December 8, 2004. This issue could potentially result in the system failing to store images when the hard disk drive capacity exceeds 170 images instead of the expected 400. Systems may also fail to boot or lose patient data as this image capacity is approached.

Event Detail

Event Id 62449
Product Type Devices
Status Terminated
Recalling Firm GE OEC Medical Systems, Inc
City Salt Lake City
State UT
Country US
Voluntary / Mandated FDA Mandated
Recall Initiation Date 2007-03-27
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US (nationwide) and the countries of: Canada, France, Japan, and China.
-
-