| Product Description | I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood. |
|---|---|
| Recall Number | Z-2133-2012 |
| Classification | Class II |
| Code Info | List number 03P76-25 Lot number S10246A |
| Product Distributed Qty | 52,750 cartridges |
| Reason For Recall | i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer. |