Product Detail

Product Description Smith & Nephew 5.5 mm Dyonics Bonecutter ,Electroblade Full Radius Part Number: 7205962 Product Usage: The Dyonics Electroblade Resector is indicated for use in arthroscopic surgical procedures of large and small articular cavities. The Dyonics Electroblade Resector is effective in soft and osseous tissue resection and in hemostasis of bleeding vessels. It is intended for arthroscopic procedures using saline solution, Ringers lactate or other conductive solutions as an irrigant under direct or video assisted fiberoptic visualization.
Recall Number Z-2145-2012
Classification Class II
Code Info Lot Numbers: 655844R, 950143R, 950144R, 950145R, 950146R
Product Distributed Qty 108 units
Reason For Recall Device labeled with an expiry period of 48 months instead of 36 months; an additional 12 months for which product and package reliability has not been validated.

Event Detail

Event Id 62617
Product Type Devices
Status Ongoing
Recalling Firm Smith & Nephew, Inc. Endoscopy Division
City Andover
State MA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-07-14
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - US Nationwide and the countries of : Australia, Canada, Korea, Turkey, UAE, and Japan.
-
-