• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 15, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description i-Stat BNP cartridges Abbott Point of Care Inc. Abbott Park, IL 60064 USA The i-Stat BNP test is an in vitro diagnostic test for the quantitative measurement of B-type Natriuertic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure.
Recall Number Z-2170-2012
Classification Class II
Code Info List number 06F30-01, 06F30-02, 03P93-25, 600-9010-25 All lots
Product Distributed Qty 1,004,200 cartridges US; 54,200 cartridges OUS
Reason For Recall Abbott Point of Care has determined that the frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.

Event Detail

Event Id 62533
Product Type Devices
Status Ongoing
Recalling Firm Abbott Point Of Care Inc.
City Princeton
State NJ
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-02-01
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Columbia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraquay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.
-
-