Enforcement Report - Week of August 15, 2012
-
Back to Previous Page
Product Detail
| Product Description | EasyLink(TM) Informatics System, Part Number 1000034805. EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-Vitro diagnostics devices. |
|---|---|
| Recall Number | Z-2172-2012 |
| Classification | Class II |
| Code Info | Version 5.0 and Version 5.0 Service Pack 1 through 4 |
| Product Distributed Qty | 237 |
| Reason For Recall | There is a potential for the EasyLink system to display and transmit to the Laboratory Information System networks an incorrect result following the EasyLink systems application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex CA-7000 or CA-1500 coagulation analyzer does not match the EasyLink system settings on the following page. |
Event Detail
| Event Id | 62448 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
| City | Newark |
| State | DE | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-07 |
| Initial Firm Notification of Consignee or Public | Letter |