Enforcement Report - Week of August 15, 2012
-
Back to Previous Page
Product Detail
| Product Description | Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments |
|---|---|
| Recall Number | Z-2173-2012 |
| Classification | Class II |
| Code Info | Version VA11B and all previously released versions. |
| Product Distributed Qty | 15 |
| Reason For Recall | The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex(R) CA-7000 or CA-1500 coagulation analyzer does not match the syngo LDM settings. |
Event Detail
| Event Id | 62403 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
| City | Newark |
| State | DE | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-07 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX. |
-