Product Detail

Product Description Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments
Recall Number Z-2173-2012
Classification Class II
Code Info Version VA11B and all previously released versions.
Product Distributed Qty 15
Reason For Recall The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex(R) CA-7000 or CA-1500 coagulation analyzer does not match the syngo LDM settings.

Event Detail

Event Id 62403
Product Type Devices
Status Ongoing
Recalling Firm Siemens Healthcare Diagnostics, Inc.
City Newark
State DE
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-07
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX.
-
-