Enforcement Report - Week of August 22, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates.
Recall Number Z-2185-2012
Classification Class I
Code Info All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0
Product Distributed Qty 102,248 units
Reason For Recall The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.

Event Detail

Event Id 55121
Product Type Devices
Status Ongoing
Recalling Firm CareFusion 303, Inc.
City San Diego
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-06-29
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution - USA (nationwide) and the countries of: South Africa and Canada.
-
-