Product Detail

Product Description Responder 5 is a nurse call communication system that connects patients with staff in a hospital. The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more. Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers
Recall Number Z-2213-2013
Classification Class II
Code Info 1) Product Code 351000 (Branch Regional Controller); 2) Product Code 352000 (Corridor Light); 3) Product Code 352020 (Domeless Controller)
Product Distributed Qty Product Code 351000: 385 units; Product Codes 352000 and 352020: 14,808 units (total)
Reason For Recall A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path. Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run, affecting nurse call communications across multiple rooms.

Event Detail

Event Id 65657
Product Type Devices
Status Ongoing
Recalling Firm Rauland-Borg Corp
City Mount Prospect
State IL
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2013-05-10
Initial Firm Notification of Consignee or Public E-Mail
Distribution Pattern Worldwide distribution - US: (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NE, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and WV., and the countries of : Australia, Canada, Columbia, Lebanon, Mexico and Saudi Arabia.
-
-