• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 22, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
Recall Number Z-2216-2012
Classification Class III
Code Info Version 6.3.0 and Later
Product Distributed Qty 103 sites
Reason For Recall In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or subtests.

Event Detail

Event Id 62344
Product Type Devices
Status Completed
Recalling Firm Sunquest Information Systems, Inc.
City Tucson
State AZ
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2011-04-28
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Distribution Pattern Nationwide Distribution
-
-