• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of August 29, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.
Recall Number Z-2224-2012
Classification Class II
Code Info Catalog Number: 32-422801 and Lot Number Identification: ZB111101, ZBll1201, ZB120101, ZB120501, ZB111102, and ZB120201.
Product Distributed Qty 119 units
Reason For Recall A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension

Event Detail

Event Id 62745
Product Type Devices
Status Ongoing
Recalling Firm Biomet U.K., Ltd.
City Bridgend, South Wales
State
Country GB
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-07-27
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.
-
-