Product Detail

Product Description Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in human serum and plasma in the assessment of thyroid function and diagnosis of thyroid disorders when T3 levels are between 0.1 and 8.0 ng/mL.
Recall Number Z-2227-2012
Classification Class III
Code Info All Lots
Product Distributed Qty 117,385 units total (33,918 units in US)
Reason For Recall A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.

Event Detail

Event Id 62791
Product Type Devices
Status Ongoing
Recalling Firm Beckman Coulter Inc.
City Brea
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2011-10-11
Initial Firm Notification of Consignee or Public Letter