Enforcement Report - Week of August 29, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description UniCel DxI 800 Access Immunoassay Systems, Part Numbers: Dxl 800: 937100; Dxl 800 Remanufactured: A25288; Dxl 800 with spot B: A71456; Dxl 800 Reconditioned: A25285. The DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Recall Number Z-2257-2012
Classification Class II
Code Info Part Numbers: Dxl 800: 937100; Dxl 800 Remanufactured: A25288; Dxl 800 with spot B: A71456; Dxl 800 Reconditioned: A25285.
Product Distributed Qty 2442 units total (784 units in US)
Reason For Recall The recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results.

Event Detail

Event Id 62706
Product Type Devices
Status Terminated
Recalling Firm Beckman Coulter Inc.
City Brea
State CA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2011-10-13
Initial Firm Notification of Consignee or Public Letter
Distribution Pattern Worldwide distribution: USA (nationwide) and countries of: Algeria, Angola, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea,Republic of; Kuwait, Liechtenstein, Malaysia, Mayotte, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Philippines, Poland, Puerto Rico, Romania, Russian Federation, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.
-
-