• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Enforcement Report - Week of September 5, 2012

  • Print
  • Share
  • E-mail
-

Product Detail

Product Description Canon Solid Stale X-ray Imaqer (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.
Recall Number Z-2265-2012
Classification Class II
Code Info Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA.
Product Distributed Qty 468 units
Reason For Recall Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

Event Detail

Event Id 62828
Product Type Devices
Status Ongoing
Recalling Firm Canon Inc.
City Tokyo
State
Country JP
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2010-11-01
Initial Firm Notification of Consignee or Public Telephone
Distribution Pattern Nationwide Distribution (USA)
-
-