Product Detail

Product Description QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.
Recall Number Z-2284-2012
Classification Class II
Code Info Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190
Product Distributed Qty 64 units
Reason For Recall STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

Event Detail

Event Id 62776
Product Type Devices
Status Ongoing
Recalling Firm Steris Corporation
City Mentor
State OH
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2012-08-10
Initial Firm Notification of Consignee or Public Visit
Distribution Pattern Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.
-
-