Enforcement Report - Week of September 12, 2012
-
Back to Previous Page
Product Detail
| Product Description | Siemens Mobilett Mira. Mobile X-Ray System. |
|---|---|
| Recall Number | Z-2322-2012 |
| Classification | Class II |
| Code Info | Model number 10273100 -- serial numbers 1034, 1038, 1041, 1042, 1043, 1053, 1057, 1070, 1079, 1081, 1088, 1098, 1099, 1102, 1103, 1105, 1106, 1107, 1108, and 1112. |
| Product Distributed Qty | 20 |
| Reason For Recall | During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to replace motor drive handles and circuit board connection cables on all affected Mobilett Mira systems. |
Event Detail
| Event Id | 63042 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Siemens Medical Solutions USA, Inc |
| City | Malvern |
| State | PA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-08-21 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI. |
-