| Product Description | SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin. |
|---|---|
| Recall Number | Z-2334-2012 |
| Classification | Class II |
| Code Info | PN 8000809705; Serial numbers 248-1298 |
| Product Distributed Qty | 1948 units |
| Reason For Recall | The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord. |
Event Detail
| Event Id | 61869 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | JK Products & Services, Inc |
| City | Jonesboro |
| State | AR | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2005-07-08 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA (nationwide) and Canada |
-