Enforcement Report - Week of September 19, 2012
-
Back to Previous Page
Product Detail
| Product Description | Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patient Interface***Medtronic Xomed, Inc., 6743 Southpoint Drive North Jacksonville, FL 32216-0980 USA***Rx Only***Lot***SN:***" Top of unit is labeled in part: "***NIM-Response 3.0***Medtronic***" Bottom/Underside of unit is labeled in part: "***5X20mm FA.L., 32mA. 250VAC***REF***MEDTRONIC XOMED INC. JACKSONVILLE, FL USA***" Cord is labeled in part: "***Ref.***Mfg Date***Sn***" The NIM 3.0 Patient Interfaces [REF #'s 8253210 (Pulse), 8253200 (Response), 8253410 (Neuro)] is an EMG Nerve Integrity Monitor intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots. |
|---|---|
| Recall Number | Z-2343-2012 |
| Classification | Class II |
| Code Info | 8253410 (Neuro). All Serial Numbers. |
| Product Distributed Qty | 53 |
| Reason For Recall | In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedure. If the delay occurs after a patient has already been anesthetized, the resulting delay could extend the patient's time under anesthesia if time is required to troubleshoot the system or if it is rendered inoperable. Models Ref 8253200 (Response) and Ref 8253410 (Neuro) were distributed in the US and Models Ref 8253200 (Response), Ref 8253210 (Pulse) and 8253410 (Neuro) were distributed internationally. |
Event Detail
| Event Id | 63008 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Medtronic Xomed, Inc. |
| City | Jacksonville |
| State | FL | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2010-09-27 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide distribution: USA (nationwide) and countries of: Australia, Austria, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Iran, Ireland, Israel, Italy, Japan, Lebanon, Libya, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudia Arabia, Singapore, Slovak, Spain, SRI Lanka, Sweden, Turkey, United Arab Emirates, and United Kingdom. |
-