Enforcement Report - Week of September 26, 2012
-
Back to Previous Page
Product Detail
| Product Description | Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine. |
|---|---|
| Recall Number | Z-2385-2012 |
| Classification | Class II |
| Code Info | Lot numbers: 019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782, 021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, and 023266. |
| Product Distributed Qty | 12,065 Sterile Distraction Screws (2413 boxes); 384 Sterile XL Screws (128 boxes) in distribution |
| Reason For Recall | This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch. |
Event Detail
| Event Id | 62899 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Tedan Surgical Innovations Llc |
| City | Sugar Land |
| State | TX | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-06-11 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide distribution: USA (nationwide) and countries of : Australia and Taiwan. |
-