Enforcement Report - Week of September 26, 2012
-
Back to Previous Page
Product Detail
| Product Description | The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole Body Computed Tomography X-ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. |
|---|---|
| Recall Number | Z-2404-2012 |
| Classification | Class III |
| Code Info | System Code #, 728246, Serial #, 6462, 6447, 6495, 6514, 6467, 6482, 6473, 6544, 6439, 6453, 6436, 6536, 6448, 6449, 6471, 6498, 6502, 6504, 6452, 6510, 6508, 6438, 6513, 6456, 6505, 6457, 6442, 6454, 6461, 6469, 6485, 6458, 6475, 6487, 6531, 6520, 6490, 6470, 6437, 6497, 6481, 6489, 6488, 6455, 6543, 6499, 6493, 6443, 6463, 6476, 6477, 6478, 6526, 6545, 6546, 6518, 6491, 6450, 6451, 6468, 6464, 6466, 6459, 6472, 6444, 6445, 6529, 6530, 6440, 6532, 6515, 6486, 6507, 6512, 6483, 6524, 6480, 6479, 6509, 6460, 6441, 6465, 6474, 6484, 6492, 6503, 6511, 6517, 6519, 6523, 6525, 6549, 6555 |
| Product Distributed Qty | 93 units |
| Reason For Recall | Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button. |
Event Detail
| Event Id | 63006 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Philips Medical Systems (Cleveland) Inc |
| City | Cleveland |
| State | OH | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-08-28 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - US Nationwide including the states of: AL, CA, GA, MA, MD, NC, OH, PA, TX, and WA; and to the following countries of: Algeria, Oman, Australia, Brazil, China, Saudi Arabia, United Arab Emirates, and India. |
-