Product Detail

Product Description Kimberly-Clark MIC PEG Kit - Pull Type The Kimberly-Clark MIC PEG Kits - Pull type are Percutaneous Endoscopic gastrostomy (PEG) tubes intended for a non-surgical "pull" endoscopic placement technique. The PEG tube is held in place during use with an external retention disk.
Recall Number Z-2416-2012
Classification Class II
Code Info Lots AW916002A, AW916102D
Product Distributed Qty 766 units
Reason For Recall The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

Event Detail

Event Id 63001
Product Type Devices
Status Completed
Recalling Firm Kimberly-Clark Corporation
City Roswell
State GA
Country US
Voluntary / Mandated Voluntary: Firm Initiated
Recall Initiation Date 2010-06-22
Initial Firm Notification of Consignee or Public Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit