Enforcement Report - Week of October 3, 2012
-
Back to Previous Page
Product Detail
| Product Description | Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation. |
|---|---|
| Recall Number | Z-2453-2012 |
| Classification | Class II |
| Code Info | Software version 5.0.1and 5.0.2 |
| Product Distributed Qty | 46 units |
| Reason For Recall | Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180°around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180° around vertical axis as well as 180° around a horizontal axis. |
Event Detail
| Event Id | 63180 |
|---|---|
| Product Type | Devices |
| Status | Ongoing |
| Recalling Firm | Hologic, Inc. |
| City | Bedford |
| State | MA | Country | US |
| Voluntary / Mandated | Voluntary: Firm Initiated |
| Recall Initiation Date | 2012-09-14 |
| Initial Firm Notification of Consignee or Public | Letter |
| Distribution Pattern | Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland. |
-