J Orthop Res 2015 Dec;33(12):1868-74

Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system.

Brock TM, Sidaginamale R, Rushton S, Nargol AV, Bowsher JG, Savisaar C, Joyce TJ, Deehan DJ, Lord JK, Langton DJ

Abstract

Taper wear at the head-neck junction is a possible cause of early failure in Large head metal-on-metal (LH-MoM) hip replacements. We hypothesized that: (i) taper wear may be more pronounced in certain product designs; and (ii) an increased abductor moment arm may be protective. The tapers of 104 explanted LH-MoM hip replacements revised for adverse reaction to metal debris (ARMD) from a single manufacturer were analyzed for linear and volumetric wear using a co-ordinate measuring machine. The mated stem was a shorter 12/14, threaded trunnion (n = 72) or a longer, smooth 11/13 trunnion (n = 32). The abductor moment arm was calculated from pre-revision radiographs. Independent predictors of linear and volumetric wear included taper angle, stem type, and the horizontal moment arm. Tapers mated with the threaded 12/14 trunnion had significantly higher rates of volumetric wear ([0.123 mm3 /yr vs. 0.402 mm3 /yr] [t = -2.145, p = 0.035]). There was a trend to larger abductor moment arms being protective (p = 0.055). Volumetric wear rate was significantly higher in the tapers mated to the shorter, threaded 12/14 trunnions. Design variation appears to play an important role in taper-trunnion junction failure. We recommend that surgeons bear these findings in mind when considering the use of a short, threaded trunnion with a cobalt-chromium head.


Category: Journal Article
PubMed ID: #26135357 DOI: 10.1002/jor.22970
Includes FDA Authors from Scientific Area(s): Medical Devices
Entry Created: 2016-02-19 Entry Last Modified: 2018-11-25